S-CATH M is the only fast-response multi-sensor esophageal temperature probe visible in 3D mapping systems that is designed to provide high-density, edge-to-edge coverage of the esophagus during therapeutic procedures performed in the operating room and laboratory. EP.
Details of the product
- Four new electrodes
Mark the center line of the probe. - Minimal metal exposure
Pebax insulation covers 75% of the electrode surface, limiting the amount of metal exposed to only what is absolutely necessary to be detected in 3D cardiac mapping systems. - Visibility in context
The visibility of impedance-based 3D cardiac mapping systems allows physicians to see the location of the esophageal tube* in the context of nearby anatomical structures. - Reliable precision
12 sensors, intelligently placed, provide the same high-density edge-to-edge coverage you expect from CIRCA Scientific. - Benefits
- Visibility in 3D context
- Visibility in impedance-based 3D cardiac mapping systems allows doctors to see the location of the esophageal tube.
- Minimal metal exposure
- S-CATH M's patented electrode design uses PEBAX insulation to cover more than 50% of the electrode surface.
- High accuracy
- The new S-CATH M, with its familiar sine shape, continues to deliver the performance you expect from CIRCA Scientific.
- Edge-to-edge coverage
- It intelligently places 12 sensors along the length and width of the esophagus, ensuring that critical temperature changes are not missed.
- Visibility in 3D context
Technical Specifications
General characteristics | |
Number of temperature sensors | 12 (+1 400 series thermistor) |
Form of the implemented matrix | Flexible S shape places an array of 12 sensors throughout the esophagus |
Temperature sensor type | thermistor |
Temperature measurement range | 0ºC to 45 C |
temperature accuracy | ±0,3ºC from 25-45ºC >± 0,4ºC from 0-24,9ºC |
Probe diameter (OD) | 10F/3,3mm |
Usable length | 65 cm (tip to connector housing) |
Sensor band material | Copper/Gold |
Sensor band length | 2,5 mm |
Exposed metal (per sensor) | None |
Probe shaft material | Polyurethane (barium loaded) with medical grade Pebax® coating |
Radiopaque | Yes |
Mapping Electrodes Type | Impedance |
Material | Gold |
Quantity | 4 |
Size | 2,5 mm |
Exposed metal (per electrode) | 15,76 mm 2 |
Radiopaque | Yes |
Connector(s) | Unidirectional connector to temperature cable, Unidirectional connector to mapping cable |
ELECTRICAL SAFETY | Complies with IEC 60601-1:2005 + A1:2012 when used with monitor and CIRCA scientific interconnect cable |
# of styluses | 1 |
Packaging unit | 1pc/box shelf |
Packaging Description | Thermofrom tray with Tyvek cover |
Sterilization | Sterile (ethylene oxide) |
Expiration | Three years from the date of manufacture. |